How can gas unmix liquids?
Investigating GC
A guided experiment in the Stemious app: students read it first, run it at the bench with the sensors listed below, and their data and conclusions are saved to their record. The first steps are below in View mode; the full lab is in the app.
Try it in View mode
Alpha: being verified by teachersThis is how the lab opens in the Stemious app: the first 5 of its 6 steps. To run it with the sensors, record data and take the quiz, students use the app.

Wonder
How do detective labs unscramble five clear liquids mixed in one tiny drop?
Show the answer
They send the vaporized liquid through an obstacle-course tube. Today, you will tune the heat and air push to watch each chemical cross the finish line one by one.

Introduction
How Can Gas Unmix Liquids?
Imagine five runners lined up on an obstacle track. Some run super fast without stopping. Others keep getting stuck in sticky mud puddles. In gas chromatography, a machine does the exact same thing with a blended liquid! It boils a tiny drop into vapor and blows it through a microscopic sticky tube. Fast runners exit first, and sticky runners exit last. Today, you will control the heat and air pressure to unscramble five mixed liquids.
Driving question: Can warm air and a sticky tube separate a tangled mixture of liquids into pure parts?
What you will learn
Today you will find out
- I can explain how a gas chromatograph separates mixed liquids.
- I can spot distinct peaks on a graph and find when each chemical exited the tube.
- I can show how changing temperature and pressure gives clearer peaks.
- I can follow safety rules for volatile chemicals and hot injection ports.
- I can name real-world careers that use gas chromatography to test unknown samples.
Why it matters: Perfume makers, airport safety agents, and environmental scientists use gas chromatography every day to identify mystery vapors.
Safety first
Scientist Rules for Today
- Wear safety goggles, lab coat, and protective gloves throughout the experiment.
- Purple gloves = teacher does it. Blue gloves = you do it.
- Methanol and solvent vapors are flammable and harmful to breathe; keep vials in a fume hood.
- The injection needle is sharp and fragile. Only the teacher handles the syringe.
- The injection port on the Mini GC gets hot (up to 120 °C). Never touch the metal nut.
- Look for the ADULT NEEDED badge. When shown, your teacher does the step.
- Wash your hands fully with soap and water after the lab.
What you need
- Go Direct Mini GC1 per station · teacher
- Tablet running Instrumental Analysis1 per group
- 1 μL glass syringe1 per station · teacher
- 5-component solvent mixture in sealed vial1 vial (2 mL) · teacher
- Methanol cleaning solvent in sealed vial1 vial (2 mL) · teacher
- Lint-free tissue box1 box
- Splash goggles and nitrile gloves1 set per person

Step 1 of 6 Teacher does this
Connect Mini GC and configure Run 1 profile
- Open Vernier Instrumental Analysis on your tablet. Connect to the Mini GC with Bluetooth or USB cable.
- Select 'Gas Chromatography' to start a new experiment.
- Set Start and Final Temperature to 120 °C, and Pressure to 18.0 kPa. Tap Apply.
- Power cord is securely plugged into mains and Mini GC.
- Vials sit safely on a tray and do not roll.
- The work area has good ventilation.
- Go Direct Mini GC is connected to 24 V power and switched on.
Careful: The internal heater draws significant current and the port will start to warm. Make sure that the power adapter is authentic Vernier 24V DC. Keep cables away from walkways.

Step 2 of 6 Teacher does this
Adult rinses the glass micro-syringe
- While the machine heats up, an adult cleans the 1 uL glass syringe with methanol.
- Dip the needle tip into the methanol vial and draw up 0.4 uL.
- Gently squirt the methanol onto a lint-free wipe.
- Repeat this flush twice so the syringe is clean and the plunger slides smoothly.
- Methanol vial is handled inside the fume hood or ventilated area.
- Goggles and gloves are securely worn.
- The delicate syringe needle is straight and undamaged.
Careful: Methanol is toxic and flammable. If you force the fragile plunger, it will bend permanently. Never pull the plunger past 50% volume. Gently expel liquid onto a wipe away from sparks.

Step 3 of 6 Teacher does this
Adult draws 0.3 microlitres of mixture
- The adult dips the syringe needle into the 5-solvent mixture vial.
- Carefully draw up exactly 0.30 uL of liquid into the glass barrel.
- Gently wipe the needle from barrel to tip with a fresh lint-free wipe.
- Check that the Mini GC LED is solid green, signalling it is heated and ready.
- Mini GC LED shows solid green.
- Sample volume is exactly 0.3 uL (not more).
- Students stand a safe step back and watch.
Careful: More than 0.3 uL will flood the column and ruin the graph with huge off-scale peaks. Read the barrel marks closely. Wipe the outside of the needle so no extra drop clings.

Step 4 of 6 Teacher does this
Inject Run 1 (120 °C, 18 kPa)
- The adult guides the needle straight into the Mini GC injection port until seated.
- Push down the plunger smoothly and completely.
- Immediately press the red start button on the Mini GC and withdraw the needle.
- Watch the graph on your tablet as data collection runs for 10 minutes.
- The needle is completely vertical and not bent.
- Student stands ready with finger near the tablet or start button.
- Injection port area is clear of loose papers.
Careful: The injection port is at 120 °C. Contact with the metal causes burns. Force on the needle bends it. Only the adult touches the syringe. Withdraw the needle straight up immediately after you push the plunger.

Step 5 of 6 Teacher does this
Run 2: Lower the pressure to 10 kPa
- On the tablet, tap the profile settings for Data Set 2.
- Keep temperature at 120 °C, but change Pressure to 10.0 kPa.
- Tap Apply. Wait for the green ready light on the Mini GC.
- The adult flushes the syringe, draws 0.3 uL of mixture, and injects for Run 2.
- Pressure is set to 10.0 kPa in the app.
- Syringe has been flushed with methanol before the new draw.
- Mini GC LED is green.
Careful: An uncleaned syringe leaves old liquid inside. This causes ghost peaks on the new graph. Always flush the syringe 3 times with methanol between injections.
1 more step, then data, observations, conclusions and a quiz
The full lab is free for teachers. Tell us where to send it and we will share it, along with the app for your students.
Where it fits in the syllabus
- Related Level AS & A Level, Chemistry, chapter 29: Organic synthesis
- Direct Level AS & A Level, Chemistry, chapter 30: Analytical chemistry
Versions and what each needs
- How can gas unmix liquids?
Organic Chemistry, 60 min, Based on a Vernier lab
Needs: Go Direct Mini GC
Vernier instruments are designed for education and are not intended for industrial, medical or commercial testing.