How fast does sugar split?
Rxn Kinetics of Sucrose
A guided experiment in the Stemious app: students read it first, run it at the bench with the sensors listed below, and their data and conclusions are saved to their record. The first steps are below in View mode; the full lab is in the app.
Try it in View mode
Alpha: being verified by teachersThis is how the lab opens in the Stemious app: the first 5 of its 7 steps. To run it with the sensors, record data and take the quiz, students use the app.

Wonder
Can a beam of twisting light reveal how fast sugar breaks apart?
Show the answer
Sugar molecules twist light waves like tiny corkscrews. As sugar breaks apart into simpler sugars, the twist changes direction. Today you will clock this chemical reaction using light.

Introduction
How Fast Does Sugar Split?
Table sugar looks simple, but its molecules are built like two rings locked together. In plain water, it takes years for those rings to come apart. But add a chemical helper, and the rings snap apart in minutes! How can we see this happen without waiting years? We shine lined-up light through the liquid. Sugar twists the light to the right. As it splits, the twist fades. Today, you will use a polarimeter to clock this reaction in real time.
Driving question: Can we use twisting light waves to watch sugar molecules split in real time?
What you will learn
Today you will find out
- I can explain how sugar molecules twist light.
- I can observe chemical helpers speeding up sugar breakdown.
- I can read a polarimeter graph to compare reaction speeds.
- I can handle a polarimeter cell safely and follow chemical safety rules.
- I can name everyday products made by splitting sugar.
Why it matters: Food scientists, pharmacists and doctors use polarimeters to check medicine purity, test honey quality, and make soft candy centres.
Safety first
Scientist Rules for Today
- Wear splash goggles, protective gloves and a long-sleeve lab coat throughout the experiment.
- Purple gloves = teacher does it. Blue gloves = you do it.
- DANGER: 6.0 M Hydrochloric acid is strongly corrosive. Only an adult handles the concentrated acid inside a fume hood.
- Never taste or drink any sugar solutions in the lab. Lab glassware is never used for food.
- Handle the glass polarimeter cell with two hands. Never touch the optical windows at the ends.
- Look for the ADULT NEEDED badge on steps where you mix acid. Stand back and observe.
- Wash your hands fully with soap and warm water when you finish.
What you need
- Go Direct Chemical Polarimeter1 per group
- Polarimeter sample cell1 per group
- Tablet running Instrumental Analysis1 per group
- Sucrose solutions (15% and 30%)100 mL each
- 6.0 M Hydrochloric acid solution10 mL per test · teacher
- Invertase enzyme solution (15 mg/mL)25 mL per group
- Distilled water wash bottle250 mL per group
- 100 mL beakers and 25 mL graduated cylinder2 beakers, 1 cylinder
- Splash goggles and protective gloves1 set per person

Step 1 of 7
Power on and pair your polarimeter
- Push the power button once to turn on the Go Direct Polarimeter.
- Open the Vernier Instrumental Analysis app on your tablet.
- Connect to your polarimeter with Bluetooth or a USB cable.
- The bench surface is clean and dry.
- All liquids are set at least 30 cm away from the tablet and sensor.
- You wear safety goggles and gloves.
- The green status light on the sensor flashes steadily.
Careful: Cables can get snagged, or spilled liquids can damage electronics. Keep cords neat. Place all bottles away from the tablet.

Step 2 of 7
Calibrate the polarimeter with distilled water
- Rinse the polarimeter sample cell with distilled water.
- Fill the cell with distilled water to a height of exactly 10.0 cm.
- Slide the cell into the polarimeter, close the lid, and tap Finish Calibration in the app.
- The outside of the glass cell is wiped dry with a lint-free tissue.
- No air bubbles stick inside the liquid column.
- The sample cell is held upright by its cylindrical body.
- Read the bottom of the meniscus at eye level.
Careful: Air bubbles or smudges on the end glass block light and cause bad calibration. Wipe the optical glass windows clean with a soft wipe and gently tap out any air bubbles.

Step 3 of 7
Measure optical rotation of pure sucrose
- Empty the water cell and rinse it with 2 mL of the 15% sucrose solution.
- Fill the cell to 10.0 cm with 15% sucrose. Put it in the polarimeter.
- Tap Start. Record the angle. Repeat the measurement with the 30% sucrose solution.
- The cell height is recorded to the nearest 0.1 cm.
- The outside of the tube is completely dry before you insert it.
- Do not taste the sugar liquid.
Careful: If you do not fill the cell fully or if you tilt it, the distance the light travels changes. Keep the tube vertical and read the liquid meniscus accurately.

Step 4 of 7 Teacher does this
Mix acid with sucrose in fume hood
- In Instrumental Analysis, select Kinetics mode, and set rate to 1 sample/s and duration to 1800 seconds.
- Your teacher pours 10 mL 30% sucrose and 10 mL 6.0 M hydrochloric acid into a 100 mL beaker.
- Your teacher swirls gently and immediately pours the mixture into the clean polarimeter sample cell to 10.0 cm.
- The teacher wears chemical splash goggles, heavy nitrile gloves, and a buttoned lab coat.
- The fume hood fan is on and pulls air correctly.
- Students stand at least 1 metre back from the fume hood.
- Students observe from a safe distance outside the fume hood sash.
- In the fume hood, carefully handle 6.0 M hydrochloric acid.
Careful: 6.0 M Hydrochloric acid produces dangerous fumes and causes severe chemical burns to skin and eyes. Only the adult handles the acid bottle. Pour slowly down the beaker wall to avoid splashes.

Step 5 of 7 Teacher does this
Collect acid reaction kinetics data
- Quickly slide the filled cell into the polarimeter and close the lid.
- Tap Start on the tablet immediately to begin the 30-minute kinetics run.
- Watch the tablet screen as the polarimeter records rotation every few seconds.
- The acid cell cap is screwed on snugly so no acid can leak inside the sensor.
- The outside of the cell is wiped completely dry with a damp wipe then dry wipe.
- Adult supervises cell insertion.
Careful: Acid liquid on the outside of the tube can corrode the polarimeter bearings and motor. Examine the tube carefully for drips before you slide it into the instrument.
2 more steps, then data, observations, conclusions and a quiz
The full lab is free for teachers. Tell us where to send it and we will share it, along with the app for your students.
Where it fits in the syllabus
- Direct Class 12, Chemistry, chapter 5: Biomolecules
Versions and what each needs
- How fast does sugar split?
Organic Chemistry, 75 min, Based on a Vernier lab
Needs: Go Direct Polarimeter
Vernier instruments are designed for education and are not intended for industrial, medical or commercial testing.