How fresh is your peroxide?

Peroxide

Grades 7-8Chemistry1 version

A guided experiment in the Stemious app: students read it first, run it at the bench with the sensors listed below, and their data and conclusions are saved to their record. The complete lab is below, free, in View mode.

The complete lab, free

Free complete lab for CBSE Class 11 Chemistry Alpha: being verified by teachers

Every screen of the lab as it appears in the Stemious app's View mode: all 8 steps, the data, observations, conclusions and quiz. Quiz answers stay in the app, so the quiz still works in class. To run it with the sensors and save each student's data, use the app.

A bottle of brown medicine-cabinet peroxide with tiny fizzing bubbles next to a water glass

Wonder

Why does an old bottle of peroxide stop fizzing on cuts?

Show the answer

Hydrogen peroxide slowly turns into plain water over time. Today you will measure its true chemical strength using an electron sensor and a purple testing liquid.

A drop of liquid fizzing with tiny white bubbles on a clean glass plate

Introduction

How Fresh Is Your Peroxide?

When you pour fresh hydrogen peroxide on a scrape, it bubbles instantly into oxygen gas and water. But if the bottle sits in the cabinet for a year, it gets weak. Inside the bottle, the peroxide slowly breaks down on its own. How can you tell if an old bottle is still good? In this lab, your teacher will drip a bright purple testing liquid into diluted peroxide. You will use an ORP sensor to catch the exact moment every peroxide molecule has reacted.

Driving question: How can we find the exact strength of medicine-cabinet peroxide using an electron sensor?

What you will learn

Today you will find out

  • I can explain how a voltage jump shows that a reaction is finished.
  • I can describe the colour change that happens during the reaction.
  • I can find the turning point on a graph of voltage versus volume.
  • I can follow safety rules when observing and recording data.
  • I can give examples of why testing chemical strength matters in daily life.

Why it matters: Pharmacists, water treatment engineers, and paper makers test chemical strengths every day so products work safely.

Large clean water filtration tanks in a municipal water treatment plant

Try These

Testing Water in City Plants

  • Drinking water plants add chlorine or ozone to kill harmful bacteria.
  • Engineers use ORP sensors to measure the water's chemical punch and ensure it is safe to drink.
Rows of amber medicine bottles being inspected on a pharmaceutical lab bench

Try These

Checking Medicine Expiry Dates

  • Liquid medicines lose their strength as molecules slowly break apart over months.
  • Quality control chemists use titration to verify that every bottle on pharmacy shelves meets safety standards.
Rolls of bright clean white paper being manufactured in a modern eco-friendly paper mill

Try These

Bleaching Paper Pulp Naturally

  • Paper factories use hydrogen peroxide to bleach wood pulp bright white without harsh chlorine.
  • Workers constantly test the peroxide concentration to use just enough and protect nearby rivers.

Warm-up question

What happens to hydrogen peroxide as a bottle sits open for many months?

  1. It slowly turns into water and oxygen gas
  2. It turns into strong acid
  3. It becomes more concentrated and dangerous
  4. Nothing changes at all

Warm-up question

What does an ORP sensor measure in a liquid mix?

  1. The electrical push from electrons being traded
  2. How hot or cold the liquid is
  3. How cloudy the liquid looks
  4. The total weight of the beaker
Diagram showing purple particles losing color as they trade tiny sparks of electricity with peroxide particles

Key idea

Chemicals trade electrons during reactions

  • Some chemical reactions involve trading tiny negative particles called electrons.
  • Permanganate grabs electrons greedily from hydrogen peroxide.
  • As long as peroxide is present, permanganate loses its purple colour instantly.
  • When all peroxide is gone, the reaction stops.

Everyday example: Batteries run flashlights because chemicals inside trade electrons to push electricity through wires.

A graph line running flat and then suddenly leaping straight upward like a cliff

Key idea

The sudden jump at the turning point

  • An ORP sensor acts like a tiny voltmeter in the liquid.
  • During the reaction, the electrical push stays fairly flat near 650 millivolts.
  • The instant the last peroxide particle reacts, the next drop of permanganate has nothing to react with.
  • This causes an immediate, steep jump in voltage up to over 1000 millivolts.

Everyday example: Like a seesaw: you can push down slowly, but once you cross the balance point, it snaps up suddenly.

Words to know: Turning point, ORP Sensor

Safety first

Laboratory Rules for Titration

  • Wear safety goggles, lab aprons, and nitrile gloves at all times.
  • Purple gloves = teacher does it. Blue gloves = you do it.
  • Sulfuric acid (4.5 M) is highly corrosive. It causes severe burns. Only the teacher measures and dispenses acid.
  • Potassium permanganate stains skin and clothes permanently and oxidises organic materials.
  • Look for the ADULT NEEDED badge. Your teacher must do those steps.
  • Keep tablets and sensor cables 30 cm away from all liquids on the bench.
  • Report any spill to your teacher immediately. Never touch liquid spills with bare hands.

What you need

  • Go Direct ORP Sensor1
  • Stir Station1
  • Electrode Support1
  • 50 mL buret1
  • 250 mL beaker1
  • magnetic stir bar1
  • Distilled water wash bottle1
  • 0.020 M potassium permanganate solution35 mL
  • 4.5 M sulfuric acid solution10 mL
  • Dilute hydrogen peroxide solution10 mL
  • Tablet running Graphical Analysis1
Teacher pouring sulfuric acid carefully into a beaker with peroxide and water

Step 1 of 8 Teacher does this

Prepare the reaction beaker

  1. Measure 10.0 mL of dilute hydrogen peroxide into a 250 mL beaker.
  2. Add 25 mL of distilled water.
  3. Carefully measure and add 10 mL of 4.5 M sulfuric acid.
Stop and check
  • Teacher wears goggles, lab apron, and purple gloves.
  • Students wear goggles and stand one pace back.
  • The teacher caps the acid bottle immediately after measurement.
  • Notice that the mixture remains completely clear and colourless.

Careful: Sulfuric acid generates heat when mixed and causes severe burns if splashed. Only the teacher dispenses the acid. Pour acid slowly down the side of the beaker.

A student placing a magnetic stir bar into the beaker on the Stir Station

Step 2 of 8

Set up the Stir Station

  1. Place the 250 mL beaker on the centre of the Stir Station platform.
  2. Carefully slide the white magnetic stir bar into the beaker.
  3. Turn the Stir Station dial so the bar spins smoothly and does not splash.
Stop and check
  • Gloves and goggles on.
  • Beaker sits flat on the Stir Station ring.
  • The Stir Station dial points to zero before you plug it in.
  • Switch on the Stir Station light to illuminate the liquid from below.

Careful: If you drop the stir bar hard, you can crack the beaker bottom. Tilt the beaker gently and let the stir bar slide down the glass wall.

Teacher pouring purple liquid through a funnel into the top of the buret

Step 3 of 8 Teacher does this

Fill the buret

  1. Attach the buret clamp and 50 mL buret to the stand.
  2. Place a waste cup beneath the buret. Rinse the buret with purple permanganate.
  3. Fill the buret above the 0.0 mL mark with 0.020 M permanganate.
Stop and check
  • Buret stopcock is closed (horizontal) before you pour.
  • Small plastic funnel is seated in top of buret.
  • You pour below eye level to prevent face splashes.
  • Drain a few drops so the tip is completely filled with liquid and free of bubbles.

Careful: Potassium permanganate spills stain tables and skin brown-purple. Pour through a small funnel slowly; wipe any drips with a damp paper towel.

A student positioning the Go Direct ORP sensor into the beaker using the electrode support

Step 4 of 8

Position the ORP sensor

  1. Slide the Go Direct ORP sensor into the Electrode Support clamp.
  2. Rinse the sensor tip fully with distilled water over a waste cup.
  3. Lower the sensor into the 250 mL beaker until the liquid covers the red tip.
Stop and check
  • Storage bottle unscrewed and set safely aside.
  • Distilled water rinse completed.
  • Stir bar spins steadily before you lower the probe.
  • Make sure that the sensor does not touch the stir bar or the buret tip.

Careful: The stir bar can strike the fragile platinum glass tip and break it. Position the probe near the side wall of the beaker, at least 1 cm above the stir bar.

Tablet running Graphical Analysis showing live millivolt meter and Volume entry dialog

Set up the sensor

Configure Graphical Analysis for Titration

Sensor
Go Direct ORP Sensor
App
Vernier Graphical Analysis
Why this mode
You record one voltage point for each millilitre of purple liquid added from the buret.
Before you start
Factory calibration stored in sensor is accurate. No calibration required.
Graphic comparing a straight diagonal line graph against an S-shaped step graph with a question mark

Predict

As purple liquid is added drop by drop, what do you predict the graph will look like?

Think about our key ideas: the reaction keeps trading electrons until the peroxide runs out.

  • The voltage will stay steady, then suddenly jump up high when the peroxide is used up
  • The voltage will climb in a perfectly straight diagonal line from start to finish
  • The voltage will drop down to zero immediately upon the first drop
  • The voltage will bounce wildly up and down without any pattern

Students record a prediction in the app before collecting data.

A student tapping Keep Point on the tablet screen while the initial point is stored

Step 5 of 8

Record the starting point (0 mL)

  1. Tap Collect on the tablet to start data collection.
  2. Before you open the buret, wait for the voltage number to become steady.
  3. Tap Keep. Type 0 for Volume (mL). Tap Keep Point.
Stop and check
  • Stir bar spins steadily.
  • Buret stopcock is completely closed.
  • Live voltage reading is stable on screen.
  • Make sure that the first dot is at zero millilitres on your graph.

Careful: If you type the wrong volume number, you create an incorrect first data point. Always make sure that you enter 0 before any liquid drops from the buret.

Adult hand adjusting buret valve while purple drops vanish in the swirling beaker

Step 6 of 8 Teacher does this

Add permanganate in 1 mL steps

  1. Carefully open the stopcock to add 1.0 mL of purple liquid.
  2. Wait 5 seconds until the millivolt reading stabilizes.
  3. Read the exact buret level, tap Keep, and type the volume.
Stop and check
  • Adult operates the buret stopcock.
  • Student watches the buret scale at eye level.
  • Stir bar continues to spin smoothly.
  • Notice that the purple drop turns completely colourless almost immediately in the liquid.

Careful: If you open the stopcock too wide, you add too much liquid at once. Turn the stopcock handle slowly to control the flow drop by drop.

Beaker with faint pink glow and tablet showing a dramatic steep vertical jump in the graph

Step 7 of 8 Teacher does this

Approach the turning point drop by drop

  1. Slow down: add permanganate drop by drop now.
  2. Watch for a single drop that leaves a persistent faint pink blush in the beaker.
  3. At this exact moment, the voltage shoots upward by hundreds of millivolts!
Stop and check
  • Stir speed is brisk so each drop mixes instantly.
  • Student watches the tablet screen closely.
  • Teacher adds one drop at a time.
  • Around 18 mL, notice the purple colour takes a few seconds longer to fade.

Careful: If you add liquid too fast, you will miss the exact millilitre value of the vertical jump. Add one drop. Wait for the voltage to settle. Record the data. Add the next drop.

Completed S-shaped titration curve displayed on the tablet screen

Step 8 of 8

Complete the curve and stop

  1. Add 2 to 3 more millilitres after the jump to establish the upper flat line.
  2. Tap Stop on the tablet screen to end data collection.
  3. Close the buret stopcock completely. Turn off the Stir Station.
Stop and check
  • Buret stopcock is completely closed.
  • Stir Station dial is turned back to 0.
  • Data collection has been stopped.
  • The voltage will level off between 1150 and 1200 millivolts.

Careful: If you leave the buret valve open, the beaker will overflow onto the bench. Double check that the stopcock handle is in the horizontal (closed) position.

Read your graph

Finding the Turning Point

  • The line stays almost flat from 0 mL to about 19 mL.
  • Between 19.5 mL and 20.5 mL, the voltage climbs steeply like a cliff.
  • The middle of that steep vertical jump is the turning point: approximately 20.1 mL.
  • This volume tells us how much peroxide was originally in the sample.

Troubleshoot

If something looks wrong

The solution turns murky brown and does not stay clear.
You did not add enough acid. Brown manganese dioxide formed. Ask your teacher to prepare a fresh sample with the full 10 mL of acid.
The voltage reading jumps wildly on the screen.
The stir bar hits the sensor or the electrode tip touches air bubbles. Adjust the clamp so the probe tip is securely under the surface.
The sensor light blinks red and will not connect.
Connect the Go Direct ORP sensor. Open the Sensor menu in Graphical Analysis. Tap the serial number of your sensor.
The graph shows a diagonal ramp instead of a steep jump.
You added permanganate too fast in big chunks. In your next trial, add drops slowly (one drop every two seconds) near 19 mL.
Rinsing the ORP sensor with distilled water and packing into storage bottle

Clean up

Clean up like a scientist

  1. The teacher pours all pink reaction mixtures into the heavy metal chemical waste container.
  2. Rinse the ORP sensor with plenty of distilled water and screw it into its pH 4 storage bottle.
  3. The teacher drains permanganate that remains in the buret. The teacher rinses it fully with tap water.
  4. Wipe the Stir Station top clean with a damp paper towel. Wash your hands with soap.

Disposal: You must neutralise manganese and acidic waste or collect it in an approved school chemical waste jug. Do not pour heavy metal oxidisers down the regular drain.

Observation

When purple permanganate first dropped into the peroxide, the purple colour vanished almost immediately.

In the app, students agree or disagree from their own data.

Observation

Between 0 mL and 18 mL, the voltage reading stayed fairly flat near 620 to 650 millivolts.

In the app, students agree or disagree from their own data.

Observation

Near 20 mL, a single drop caused the liquid to keep a faint pink colour and the voltage leaped upward steeply.

In the app, students agree or disagree from their own data.

Observation

After the steep jump, adding more purple liquid made the voltage climb straight to infinity.

In the app, students agree or disagree from their own data.

Conclusion

Purple drops turn clear only while there is peroxide available to trade electrons with them.

Conclusion

The sudden steep jump in millivolts marks the turning point where every peroxide molecule has reacted.

Conclusion

The exact volume of purple liquid needed to trigger the voltage jump tells us the strength of the peroxide.

Think back

Think like a scientist

  • Why was adding liquid drop by drop near the turning point so important?
  • What visual sign warned you that the voltage jump was about to happen?
  • How could this method tell if a bottle of peroxide in a hospital is too old to use?

Question 1

What does the Go Direct ORP sensor measure during the titration?

  1. The push of electrons traded in the liquid, measured in millivolts
  2. The temperature increase in degrees Celsius
  3. The total weight of liquid in grams
  4. The brightness of the room lights

Question 2

What colour change told you that all the peroxide had completely reacted?

  1. A permanent pale pink colour remained in the liquid
  2. The liquid turned dark green
  3. The liquid turned milky white
  4. The liquid stayed boiling hot

Question 3

On your titration curve, what feature marks the turning point of the reaction?

  1. The middle of the steep vertical cliff in voltage
  2. The very first data point at 0 mL
  3. The lowest dip in the line
  4. Where the line crosses the x-axis

Question 4

Why did the purple drops turn clear when added during the first 15 mL?

  1. Permanganate traded electrons with peroxide, destroying the purple colour
  2. The water in the beaker simply diluted the purple dye away
  3. The stir bar absorbed the purple dye like a sponge
  4. The sensor lamp bleached the liquid with light

Question 5

Why did the voltage suddenly shoot upward near 20 mL?

  1. All peroxide was used up, so unreacted oxidising liquid built up
  2. The battery in the sensor suddenly died
  3. The liquid reached boiling temperature
  4. The stir bar started spinning backwards

Question 6

If an old, degraded bottle of peroxide is tested, how will its turning point volume compare to fresh peroxide?

  1. It will need LESS purple liquid to reach the turning point
  2. It will need MORE purple liquid to reach the turning point
  3. It will need exactly the same amount of purple liquid
  4. The graph will never show any voltage jump at all

Question 7

Why must 4.5 M sulfuric acid only be handled and measured by an adult?

  1. It is highly corrosive, generates heat, and causes severe skin and eye burns
  2. It stains tables purple permanently
  3. It smells sweet and might be mistaken for fruit juice
  4. It is too cold to touch safely

Question 8

What could happen if the ORP sensor touches the spinning magnetic stir bar?

  1. The spinning bar can crack the delicate platinum glass electrode tip
  2. The stir bar will turn completely purple
  3. The tablet screen will switch off
  4. The liquid will instantly freeze solid

Question 9

Why did we switch from adding 1 mL at a time to single drops near 18 mL?

  1. To find the exact volume where the vertical jump happens without overshooting
  2. Because the buret was almost empty
  3. Because the stir bar stopped spinning
  4. To give the tablet time to recharge

Question 10

Which of these real-world facilities relies on ORP sensors to ensure safety?

  1. City drinking water plants testing disinfection punch
  2. Car repair garages checking tyre air pressure
  3. Bakeries weighing flour for bread
  4. Weather stations measuring wind speed

Question 11

If your solution turns muddy brown instead of staying clear, what likely went wrong?

  1. Not enough sulfuric acid was added to support the clean reaction
  2. The beaker was kept under too much light
  3. The stir bar was spinning too fast
  4. The tablet Bluetooth disconnected

Question 12

What is the correct way to store the Go Direct ORP sensor after cleaning it?

  1. Rinse with distilled water and seal in its bottle filled with storage solution
  2. Dry the tip with paper towels and leave it lying open on the table
  3. Submerge the entire sensor body inside a beaker of acid
  4. Store it inside the freezer
Two brown medicine bottles side by side with buret apparatus in background

Try These

Fresh vs. Opened Peroxide

  • Compare a brand new, sealed bottle of 3% peroxide against one opened six months ago.
  • Titrate equal samples of both. Does the older bottle reach the turning point with less purple liquid?
  • Calculate the difference in strength to find out how quickly household peroxide expires.
Tablet screen displaying a tall sharp derivative spike pinpointing turning point

Try These

The First Derivative Peak

  • Open Graphical Analysis and add a Calculated Column for 1st Derivative.
  • Watch the S-curve transform into a sharp mountain peak.
  • The very tip of that spike points directly to the exact turning point drop!
Go Direct Drop Counter mounted on Stir Station with reagent reservoir

Try These

Automating with a Drop Counter

  • Mount the Go Direct Drop Counter beneath a plastic reagent reservoir.
  • Let purple drops fall steadily through the optical sensor beam.
  • Watch the tablet plot every single drop automatically without any manual typing!
Split scene showing peroxide bottle in a sunny window and another in a refrigerator

Challenge

Does Warmth Spoil Peroxide Faster?

  • Store one sealed sample of peroxide in a cool refrigerator and another in a warm sunlit spot for two weeks.
  • Titrate both with your teacher to measure which sample lost more of its strength.
  • Why are medicine bottles always dark amber and labelled 'store in a cool place'?

Run this lab with your class

Students follow the same screens in the Stemious app, record live sensor data and take the quiz, and their work is saved to their record.