Experiments › Environmental Science
How much nitrate is in our lake?
Nitrate in Freshwater Systems
A guided experiment in the Stemious app: students read it first, run it at the bench with the sensors listed below, and their data and conclusions are saved to their record. The first steps are below in View mode; the full lab is in the app.
Try it in View mode
Alpha: being verified by teachersThis is how the lab opens in the Stemious app: the first 5 of its 7 steps. To run it with the sensors, record data and take the quiz, students use the app.

Wonder
Why does a clear lake suddenly turn thick green like pea soup?
Show the answer
Fertilizers wash into the water and trigger massive algae blooms. Scientists call these nutrients nitrates, and today you will measure them.

Introduction
Why Do Lakes Turn Green?
Plants need food to grow, especially nitrogen. Farmers spray fertilizers on fields to boost crops. But when heavy rain falls, extra nitrogen washes off the soil into local streams. This invisible chemical is called nitrate. In water, it acts like super-fuel for microscopic algae. In days, a sparkling lake can turn into a smelly, soupy green mat that chokes out fish. Today you will act as an environmental detective to measure nitrate before disaster strikes.
Driving question: Why do some lakes suddenly turn green, and how do we measure the invisible chemical causing it?
What you will learn
Today you will find out
- I can explain where nitrates come from and how they enter water.
- I can explain why too much fertilizer makes lakes turn green and lose oxygen.
- I can soak, calibrate, rinse, and use a nitrate electrode properly.
- I can measure nitrate levels and tell if water is clean or polluted.
- I can judge whether a stream has healthy nutrient levels.
Why it matters: Clean drinking water keeps communities safe, and healthy streams protect fish and wildlife.
Safety first
Scientist Rules for Today
- Wear safety goggles, lab aprons, and nitrile gloves at all times.
- Purple gloves = teacher does it. Blue gloves = you do it.
- Never taste or drink water samples from ponds, streams, or lab beakers.
- The sensor tip is fragile and expensive: never push it against the glass bottom.
- Keep all beakers of liquid at least 30 cm away from your tablet or laptop.
- Rinse the probe tip with distilled water over a waste beaker between every test.
- Wash your hands fully with soap and warm water when you finish lab work.
What you need
- Go Direct Nitrate ISE1 per group
- Tablet running Graphical Analysis1 per group
- High Standard (100 mg/L NO3–-N)30 mL in 100 mL beaker · teacher
- Low Standard (1 mg/L NO3–-N)30 mL in 100 mL beaker · teacher
- Water sample (stream or pond)100 mL in 150 mL beaker
- Distilled water wash bottle1 bottle (250 mL)
- 250 mL waste beaker1 per group
- Lint-free tissue wipes1 box per station
- Goggles and nitrile gloves1 set per person

Step 1 of 7 Teacher does this
Confirm the 30-minute sensor soak
- Make sure that your Nitrate ISE was in High Standard (100 mg/L) for at least 30 minutes.
- Make sure that the threaded cap holds the probe tip above the bottom of the bottle.
- Look closely through the clear bottle. Make sure that liquid fully covers the small white dots near the tip.
- The electrode stands upright in its storage bottle.
- An adult confirms that the probe was in solution for at least 30 minutes.
- Table surface is clean and dry.
- Make sure that no trapped air bubbles stick to the bottom of the electrode tip.
Careful: A dry membrane causes unstable numbers that never settle down. Always keep the tip soaked in standard solution before calibration.

Step 2 of 7
Power on and pair your sensor
- Press the power button on your Go Direct Nitrate ISE once.
- Open Graphical Analysis on your tablet or computer.
- Tap Sensor Data Collection. Find your probe ID. Tap Connect.
- Tablet is placed at least 30 cm away from all liquid beakers.
- Hands and gloves are dry before you touch the screen.
- The Bluetooth light on the probe flashes green once connected.
Careful: If you tap the wrong sensor ID, your tablet connects to your classmate's probe. Match the 8-digit serial number on your probe label to the screen.

Step 3 of 7 Teacher does this
Calibrate Point 1 in High Standard
- Carefully unscrew the cap, remove the probe from its bottle, and place the tip into High Standard (100 mg/L NO3-N).
- In Graphical Analysis, tap the sensor meter in the bottom corner and tap Calibrate.
- Type 100 for Value 1. Tap Keep when the voltage reading stays steady.
- Electrode tip does not hit the glass bottom of the beaker.
- High Standard beaker is clearly labelled 100 mg/L.
Careful: If you type 1 instead of 100, you reverse the calibration curve. Double-check that you typed 100 for the High Standard before you tap Keep.

Step 4 of 7
Rinse and blot the probe tip
- Lift the Nitrate ISE carefully out of the High Standard beaker.
- Hold the probe over your waste beaker. Rinse the tip fully with distilled water.
- Gently blot the tip dry with a clean lab tissue. Never rub or scrape the flat membrane!
- Waste beaker is directly beneath the probe tip.
- You use a soft lab tissue, not rough paper towels.
Careful: Leftover High Standard drops will contaminate your Low Standard beaker. Rinse generously with distilled water and gently blot off droplets.

Step 5 of 7 Teacher does this
Calibrate Point 2 in Low Standard
- Place the tip of the ISE into the Low Standard (1 mg/L NO3-N) beaker.
- Swirl gently once, then hold still for 60 seconds as the reading settles.
- Type 1 for Value 2. When the voltage is stable, tap Keep and then tap Apply.
- Low Standard beaker contains clear 1 mg/L solution.
- You fully rinsed and blotted dry the probe before it entered the solution.
- Make sure the white reference dots are covered, but do not touch the bottom.
Careful: If you tap Keep too early, you get an inaccurate calibration curve. Wait a full 60 seconds until the voltage is stable.
2 more steps, then data, observations, conclusions and a quiz
The full lab is free for teachers. Tell us where to send it and we will share it, along with the app for your students.
Where it fits in the syllabus
- Related Level DP, Environmental Systems and Societies, chapter 4: Water and aquatic food production systems and societies
Versions and what each needs
- How much nitrate is in our lake?
Water Quality, 60 min, Based on a Vernier lab
Needs: Go Direct Nitrate Ion-Selective Electrode
Vernier instruments are designed for education and are not intended for industrial, medical or commercial testing.