Experiments › Biology
How much protein is inside?
Macromolecules and Protein Testing
A guided experiment in the Stemious app: students read it first, run it at the bench with the sensors listed below, and their data and conclusions are saved to their record. The first steps are below in View mode; the full lab is in the app.
Try it in View mode
Alpha: being verified by teachersThis is how the lab opens in the Stemious app: the first 5 of its 10 steps. To run it with the sensors, record data and take the quiz, students use the app.

Wonder
How do scientists prove how much protein is inside a shake without drinking it?
Show the answer
Special indicator dyes change colour when they grab protein chains. Today you use a light sensor to measure true protein content.

Introduction
Unlocking the Protein Secret
Look at the back of a carton of milk or a fitness shake. The label claims it has lots of protein. Protein is a giant macromolecule that builds muscles, hair, and organs. But how do food labs test that claim? Today you will use a special dye called Bradford reagent. In water, it is reddish-brown. When it hugs a protein chain, it shifts to bright blue! The darker the blue, the more protein is inside.
Driving question: How can shining a light through a drink reveal its secret protein content?
What you will learn
Today you will discover
- I can explain what proteins are made of.
- I can build and measure a protein standard curve with a light sensor.
- I can find the unknown protein amount in common drinks.
- I can tell the difference between tests for whole proteins and single building blocks.
- I can handle lab indicators safely while using micropipettes.
Why it matters: Food safety inspectors, doctors, and sports nutritionists test protein levels every day to ensure diets are healthy and labels are honest.
Safety first
Laboratory Safety Rules
- Wear safety goggles, lab coat, and nitrile gloves at all times.
- Purple gloves = teacher does it. Blue gloves = you do it.
- Quick Start Bradford Reagent contains dilute phosphoric acid and methanol. The teacher dispenses it.
- Biuret reagent contains sodium hydroxide, which can irritate skin and eyes.
- Never taste or drink any liquid in the lab. Do not drink the milk or protein drinks.
- Keep all liquids at least 30 cm away from tablets and electronic sensor ports.
- Use the chemical bins to discard all micropipette tips and chemical waste.
What you need
- Go Direct Colorimeter1
- Tablet running Graphical Analysis1
- Plastic cuvette (4 clear sides)10 with caps
- Cuvette Rack1
- 20–200 µL micropipette1
- 100–1000 µL micropipette1
- micropipette tips in rack1 box each (200 uL and 1000 uL)
- Quick Start Bradford Reagent15 mL in centrifuge tube · teacher
- Phosphate buffered saline (PBS)10 mL in centrifuge tube
- Bovine γ-globulin standard set7 tubes (0.125 to 2.0 mg/mL)
- Diluted milk and protein drink (1:50)2 microtubes (labelled M and HP)
- Lint-free tissue box1 box
- Splash goggles and nitrile gloves1 set per person

Step 1 of 10 Teacher does this
Dispense Bradford dye into cuvettes
- Place 8 clean cuvettes into your rack.
- Label one cuvette 'Blank' and the other seven with the numbers: 0.125, 0.25, 0.5, 0.75, 1.0, 1.5, and 2.0.
- An adult or supervised student pipettes 1.0 mL (1000 uL) of Bradford reagent into each cuvette.
- Check that each cuvette holds an identical level of reddish dye.
- Everyone has goggles and nitrile gloves on.
- Bradford reagent tube is kept upright in a rack.
- Pipette tips are fresh and clean.
Careful: Splashes of Bradford dye can stain skin and clothes or irritate eyes because of dilute acid. Depress the pipette plunger smoothly. Dispense slowly along the inner wall of each cuvette.

Step 2 of 10
Prepare the blank and standard ladder
- To the 'Blank' cuvette, add 20 uL of clear PBS buffer. Cap and gently invert three times.
- To each numbered cuvette, add 20 uL of the correct protein standard (0.125 up to 2.0 mg/mL).
- Always put on a new pipette tip between different tubes to prevent cross contamination.
- Cap each cuvette. Gently turn it upside-down three times to mix fully.
- Make sure that caps are tight before you invert the cuvette.
- Check tip rack: use a brand new tip for every single standard tube.
Careful: If you reuse a pipette tip, you carry high-concentration protein into low-concentration tubes and spoil the curve. Eject the plastic tip into waste immediately after you dispense each standard.

Step 3 of 10
Prepare the drink unknown cuvettes
- Take two more clean cuvettes. Label one 'M' for milk and one 'HP' for protein drink.
- Add 1.0 mL of Bradford reagent to each cuvette.
- Add 20 uL of diluted milk (tube M) to cuvette M. Cap and invert three times.
- Add 20 uL of diluted protein drink (tube HP) to cuvette HP. Cap and invert three times.
- Use pre-diluted samples (1:50) prepared by the teacher.
- Wear gloves and keep caps snug.
Careful: Undiluted milk makes the colour too dark for the sensor to read accurately. Double-check that you pipette from the 1:50 diluted stock microtubes.

Step 4 of 10
Calibrate on the blank cuvette
- Place the Blank cuvette (Bradford dye + PBS buffer) into the Go Direct Colorimeter slot and shut the lid completely.
- Press the arrow buttons on the Colorimeter until you select the 635 nm (Red) light.
- Press and hold the CAL button on the sensor, and release it when the indicator light blinks.
- The outside of the cuvette is completely dry and wiped free of fingerprints.
- The sensor lid closes completely flat without force.
- Wipe all four smooth sides clean with a lint-free tissue.
- Check that the light no longer blinks and the screen reads 0.000 absorbance.
Careful: Fingerprints or droplets on the cuvette make the blank falsely dark and corrupt all future readings. Hold the cuvette only by the upper frosted ridges. Wipe the clear sides fully.

Step 5 of 10
Measure the protein standard ladder
- In Graphical Analysis, tap 'Collect' to start to record data pairs.
- Start with cuvette 0.125. Wipe the outside with a tissue. Put it into the sensor. Close the lid.
- Wait 5 seconds until the absorbance reading stabilizes. Tap 'Keep'.
- Type '0.125' as the concentration. Tap 'OK'.
- Repeat this exact process for 0.25, 0.5, 0.75, 1.0, 1.5, and 2.0 mg/mL.
- Wipe each cuvette right before you put it into the sensor.
- Keep sensor lid firmly shut during each reading.
Careful: If you type the wrong concentration number, the software plots the point in the wrong spot on your curve. Double-check the cuvette label before you press Keep and enter the value.
5 more steps, then data, observations, conclusions and a quiz
The full lab is free for teachers. Tell us where to send it and we will share it, along with the app for your students.
Where it fits in the syllabus
- Related Class 11, Biology, chapter 9: Biomolecules
- Related Level AS & A Level, Biology, chapter 2: Biological molecules
- Direct Level DP, Biology, chapter B1.2: Proteins
Versions and what each needs
- How much protein is inside?
Advanced Biology, 75 min, Based on a Vernier lab
Needs: Go Direct SpectroVis Plus Spectrophotometer, Go Direct Colorimeter