Why is river water murky?
Turbidity and Water Quality
A guided experiment in the Stemious app: students read it first, run it at the bench with the sensors listed below, and their data and conclusions are saved to their record. The first steps are below in View mode; the full lab is in the app.
Try it in View mode
Alpha: being verified by teachersThis is how the lab opens in the Stemious app: the first 5 of its 6 steps. To run it with the sensors, record data and take the quiz, students use the app.

Wonder
Why do rushing streams turn muddy brown after a thunderstorm?
Show the answer
Rain washes loose dirt, clay, and street dust straight into rivers. Today you will measure this cloudiness using scattered light!

Introduction
How Clear Is the River?
Have you ever looked at a river after a heavy storm? The water often looks like chocolate milk. Heavy rain washes loose soil, road grime, and decayed plants into the stream. Scientists call this cloudiness turbidity. High turbidity blocks sunlight, warms the water, and chokes aquatic life. Today, you will use a digital turbidity sensor to discover exactly how clear your water is.
Driving question: Why does a clear stream turn chocolate brown after rain, and how do scientists measure the mud?
What you will learn
Today you will find out
- I can explain what causes water to look cloudy.
- I can calibrate and use a turbidity sensor correctly.
- I can measure turbidity and compare it to water quality ranges.
- I can identify why a river or pond becomes muddy after rain.
- I can describe how high turbidity harms fish and river plants.
Why it matters: Drinking water plants, wildlife rangers, and river scientists monitor turbidity every day to keep drinking water safe and rivers healthy.
Safety first
Lab Rules for Water Testing
- Wear safety goggles and lab gloves throughout the experiment.
- Purple gloves = teacher does it. Blue gloves = you do it.
- Never taste or drink any lab water samples, even if they look clear.
- Keep water bottles and cuvettes at least 20 cm away from the tablet.
- Wipe cuvettes gently with soft lint-free wipes so you do not scratch the optical surface.
- Wash your hands with warm water and soap after you complete the lab.
What you need
- Go Direct Turbidity Sensor1 per group
- Tablet running Graphical Analysis1 per group
- Plastic cuvette (4 clear sides)1 per group
- Turbidity Standard (100 NTU)1 per group
- Distilled water wash bottle1 per group
- Lint-free tissue box1 box per bench
- Sampling Chamber (250 mL / 2 L)1 bottle per group
- 100 mL beakers1 per group

Step 1 of 6
Power on and warm up the sensor
- Press the power button on the Go Direct Turbidity Sensor.
- Open Graphical Analysis on your tablet. Connect the sensor with Bluetooth.
- Leave the sensor on a flat, sturdy desk for at least 5 minutes so the internal lamp warms up.
- Desk surface is dry and steady.
- The sensor battery light is active.
Careful: Sensor movement during warm-up can cause unstable readings. Keep the sensor flat and still on the tabletop.

Step 2 of 6
Calibrate reading 1: 0 NTU blank
- Fill the cuvette with distilled water until the bottom curve touches the white mark. Cap the cuvette. Wipe the outside with a lint-free wipe.
- Align the white line with the sensor notch. Put the cuvette into the sensor.
- Close the sensor lid. Tap Calibrate in Graphical Analysis. Type 0 NTU. Tap Keep.
- The cuvette outside is completely dry and free of fingerprints.
- The cuvette is filled exactly to the line.
- Rinse the empty cuvette with distilled water.
- Wipe the cuvette outside gently with a lint-free wipe and tap out tiny air bubbles.
Careful: Fingerprints, scratches, or bubbles will scatter extra light and ruin the zero reading. Wipe the cuvette with lint-free tissue. Tap gently on the desk pad to release bubbles.

Step 3 of 6
Calibrate reading 2: 100 NTU standard
- Gently invert the sealed 100 NTU Turbidity Standard cuvette 4 times. Do not shake the cuvette. Wipe the outside dry with a lint-free wipe.
- Align the marks. Put the cuvette into the sensor. Close the lid.
- Type 100 NTU for Reading 2. Wait for a steady voltage. Tap Keep. Tap Apply.
- Standard cuvette was inverted gently, not shaken.
- The sensor lid is closed securely.
- Remove the distilled water cuvette and empty it into your waste beaker.
- Do not shake the standard cuvette.
- Wipe the sides with a lint-free tissue.
Careful: If you shake the cuvette, you introduce thousands of micro-bubbles that mimic particles and falsify calibration. Invert upside-down and right-side-up slowly four times only.

Step 4 of 6
Mix and prepare your water sample
- Take your water sample bottle and gently invert it 4 times.
- Fill the cuvette with sample water until the meniscus touches the top of the white line.
- Cap the cuvette and wipe the outside dry with a lint-free wipe.
- Sample was gently inverted, not violently shaken.
- Cuvette was rinsed first with sample water.
- Do not shake the bottle, because bubbles create false particle readings.
- Rinse your empty test cuvette with a splash of the sample water. Dump the water into the waste beaker.
Careful: If the sample sits still, dirt settles to the bottom and causes a deceptively low reading. Invert the jar immediately before you fill the cuvette.

Step 5 of 6
Measure trial 1
- Align the white marks. Put the sample cuvette into the turbidity sensor.
- Close the sensor lid fully. Tap Collect in Graphical Analysis.
- Tap Keep. Round your reading to the nearest whole NTU. Record the reading in your data table.
- Sensor lid is fully shut.
- Nobody leans on or shakes the table.
- Hold still and do not bump the table during the 10-second sample period.
Careful: If you bump the table, liquid sloshes and changes the light path while the sensor calculates the average. Keep hands off the desk while the sensor takes its 10-second average.
1 more step, then data, observations, conclusions and a quiz
The full lab is free for teachers. Tell us where to send it and we will share it, along with the app for your students.
Where it fits in the syllabus
- Direct Class 9, Science, chapter 2: Is Matter Around Us Pure?
- Related Class 10, Science, chapter 4: Carbon and its Compounds
- Related Class 8, Chemistry, chapter 11: Dissolution of Salts in Water
- Related Level DP, Environmental Systems and Societies, chapter 4: Water and aquatic food production systems and societies
Versions and what each needs
- Why is river water murky? (Earth Science)
Earth Science, 50 min, Based on a Vernier lab
Needs: Go Direct Turbidity Sensor - Why is river water murky? (Water Quality)
Water Quality, 45 min, Based on a Vernier lab
Needs: Go Direct Turbidity Sensor
Vernier instruments are designed for education and are not intended for industrial, medical or commercial testing.