Sensors › Environment and water quality
Go Direct Potassium Ion-Selective Electrode
Potassium ion concentration in water and soil extracts
An ion-selective electrode set for measuring potassium (K+) concentration in water samples. It is a Go Direct ISE Amplifier joined to a BNC potassium electrode: the amplifier reads the electrode potential in mV and converts it to mg/L, sending data by Bluetooth or USB. The gel-filled combination electrode needs no refilling, the PVC membrane module can be replaced when it ages, and the same amplifier accepts other Go Direct BNC ion electrodes.
Type: wireless ISE (Go Direct ISE Amplifier with Potassium ISE BNC electrode)
What students do with it
- Measure potassium levels in stream, soil-extract or fertiliser-runoff samples
- Carry out a two-point calibration with high and low standards
- Relate electrode potential to concentration using the Nernst relationship
- Compare water quality across sites or over time
What it records
- Potassium (mg/L as K⁺)
- mV (raw electrode potential)
What is in the box
- Go Direct ISE Amplifier connected to a Go Direct Potassium BNC electrode
- Micro USB cable
- 30 mL High Standard solution with SDS (1000 mg/L K+)
- 30 mL Low Standard solution with SDS (10 mg/L K+)
- Short-Term ISE Soaking Bottle
What else you need
- Graphical Analysis app (free) on a computer, Chromebook, iOS or Android device, connected by Bluetooth or USB
- Alternatively LabQuest 3 or LabQuest 2 (original LabQuest is not compatible)
- Distilled water for rinsing; standards for calibration (two supplied)
- Optional ionic strength adjuster (1 M NaCl) for better accuracy at low concentrations
- Also supported from Python, JavaScript (Web Bluetooth browser) and NI LabVIEW
Accessories and replacement parts
- Standard High K ISE Solution K-HST
- Standard Low K ISE Solution K-LST
- Potassium Replacement Module K-MOD
- Go Direct Potassium ISE BNC (electrode only) GDX-K-BNC
- Go Direct ISE Amplifier GDX-ISEA
- Electrode Storage Bottles, pack of 5 BTL-ES
- Go Direct Charge Station GDX-CRG
- Micro USB Cable CB-USB-MICRO
- Go Direct 300 mAh Replacement Battery GDX-BAT-300
- Go Direct Nitrate ISE BNC GDX-NO3-BNC
- Go Direct Ammonium ISE BNC GDX-NH4-BNC
- Go Direct Chloride ISE BNC GDX-CL-BNC
- Go Direct Calcium ISE BNC GDX-CA-BNC
Care and limits
- Soak the tip in High Standard for 30 minutes before use, then do a two-point calibration; do not soak longer than 24 hours
- For storage beyond 24 hours, rinse, blot dry and keep in the storage bottle with a sponge moistened with distilled water
- The membrane module has a limited life and degrades on the shelf, so buy replacements only when needed
- Never immerse above the BNC junction; the amplifier is not water resistant. Designed for educational use; not intended for industrial, medical or commercial applications
Specifications are as published by Vernier, checked on 18 November 2026. Descriptions are written by Stemious. Vernier products are designed for education and are not for medical, industrial or research-grade commercial use. Vernier and Go Direct are trademarks of Vernier Science Education.
With Stemious in India
School experiments that use it
No school experiment in the catalogue uses it yet.